Hazardous Substances | Example | |
Screening Analysis | Detailed analysis | |
Hg | X-ray spectroscopy | AAS |
Pb | ||
Cd | ||
Cr6+ | X-ray spectroscopy(Analysis of total Cr) | Ion chromatography |
PBBs / PBDEs | X-ray spectroscopy(Analysis of total Br) | GC-MS |
Sample | Given Value | Actual Value (XD-8010) |
Blank | 0 | 0 |
Sample 1 | 97.3 | 97.4 |
Sample 2 | 288 | 309.8 |
Sample 3 | 1122 | 1107.6 |
Sample | Given Value | Actual Value (XD-8010) |
Blank | 0 | 0 |
Sample 1 | 90 | 89.7 |
Sample 2 | 280 | 281.3 |
Sample 3 | 1116 | 1114.1 |
Sample | Given Value | Actual Value (XD-8010) |
Blank | 0 | 0 |
Sample 1 | 8.7 | 9.8 |
Sample 2 | 26.7 | 23.8 |
Sample 3 | 107 | 107.5 |
Sample | Given Value | Actual Value (XD-8010) |
Blank | 0 | 0 |
Sample 1 | 91.5 | 87.5 |
Sample 2 | 271 | 283.5 |
Sample 3 | 1096 | 1089.5 |
Sample | Given Value | Actual Value (XD-8010) |
Blank | 0 | 0 |
Sample 1 | 93.1 | 91.4 |
Sample 2 | 276 | 283.9 |
Sample 3 | 1122 | 1120.3 |
Cr | Br | Cd | Hg | Pb | |
1 | 1128.7 | 1118.9 | 110.4 | 1079.5 | 1109.4 |
2 | 1126.2 | 1119.5 | 110.8 | 1072.4 | 1131.8 |
3 | 1111.5 | 1115.5 | 115.8 | 1068.9 | 1099.5 |
4 | 1122.1 | 1119.9 | 110.3 | 1086.0 | 1103.0 |
5 | 1115.6 | 1123.6 | 103.9 | 1080.7 | 1114.8 |
6 | 1136.6 | 1113.2 | 101.2 | 1068.8 | 1103.6 |
7 | 1129.5 | 1112.4 | 105.3 | 1079.0 | 1108.0 |
Average | 1124.3 | 1117.6 | 108.2 | 1076.5 | 1110.0 |
Standard deviation | 8.61 | 4.03 | 4.99 | 6.54 | 10.82 |
RSD | 0.77% | 0.36% | 4.62% | 0.61% | 0.98% |
Model | NB-8010 | |
Analysis principle | Energy dispersive X-ray fluorescence analysis | |
Elements Range | S (16) – U (92) any element | |
Sample | Plastic / metal / film / solid / liquid / powder, etc., any size and irregular shape | |
X - ray tube | Target | Mo |
Tube voltage | (5-50)kV | |
Tube current | (10-1000)uA | |
Sample irradiation diameter | F1mm-F7mm | |
Filter | 15 sets of composite filter is automatically selected, and the automatic conversion | |
Detector | Imports from the United States Si-PIN detector | |
The data processing circuit board | Imports from the United States, with the use of Si-PIN detector sets | |
Sample observation | With 300,000-pixel CCD camera | |
The sample chamber size | 490 (L)´430 (W)´150 (H) | |
Analysis method | Linear linear, quadratic Code lines, strength and concentration calibration correction | |
Operating system software | Windows XP, Windows7 | |
Data management | Excel data management, test reports, PDF / Excel format saved | |
Working environment | Temperature: £30°C. Humidity£70% | |
Weight | 55kg | |
Dimensions | 550´450´395 | |
Power supply | AC220V±10%,50/60Hz | |
Determination conditions | Atmospheric environment |