Ion Chromatographic Pump | Type | High-pressure and low-pulse two-piston tandem advection pump |
Maximum Pressure | 35MPa(PEEK) |
Pressure Display Accuracy | ≤0.1MPa |
Flow Stability | (0.2-0.5)mL/min≤3%; (0.5-1.0)mL/min≤2%; >1.0mL/min≤2%; |
Allowable Deviation of Flow | (0.2-0.5)mL/min , ±5%; (0.5-1.0)mL/min, ±3%; >1.0mL/min, ±2%; |
Flow Range | 0.001~9.999mL/min |
Pressure Pulse | ≤0.5% |
Flow Precision | ≤0.1% |
Flow Accuracy | ≤0.1% |
Manual Sample Injector | Maximum Pressure | 35MPa |
External Diameter of the Tube | 1/16" |
Contact Material of the Rotor | PEEK |
Contact Material of Medium | PEEK/Ceramics |
Column Heater | Operating Temperature Range | 20~60℃(68~140℉) |
Controlling Temperature Accuracy | ±0.01℃ |
Allowable Deviation of Column Heater’s Temperature | ±1℃ |
Temperature Stability | ≤0.05℃/h |
Conduction Detection System | Type | Constant temperature auto-range conductivity detector |
Cell Volume | ≤0.8μL |
Detection Range | 0~35000μS/cm |
Detection Resolution | ≤0.0020nS/cm |
Output Voltage | -6000+6000mv(adjustable) |
Electronic Noise | 0.02nS |
Baseline Noise | ≤0.001μS/cm |
Baseline Drift | ≤0.02μS |
Operating Temperature Range | Room temperature +5℃~60℃(41 ~140℉) |
Controlling Temperature Accuracy | ±0.01℃ |
Temperature Compensation | 1.7 %/℃ |
Maximum Pressure | 10.0MPa |
Linear Range | ≥103 |
Instrument Linearity | ≥0.999 |
Quantitative Repeatability | ≤1.0% |
Qualitative Repeatability | ≤0.5% |
Minimum Detectable Concentration | Cl-≤0.0005ug/mL; Li+≤0.001ug/mL; BrO3-≤0.001ug/mL; |
Flow System | Plastic Flow Path | Made of PEEK materials |
Six-way Valve | PEEK material, pressure 5000psi; Independent automatic collecting and flow function. |
Suppressor | Type | Self-Regenerating electrolytic micro-membrane suppressor |
Maximum Pressure | 6.0MPa |
Dead Volume | <50μL |
Power Supply | 150W |
External Size(L*W*H*) | 325*400*500 |
Net Weight(KG) | 21 |
Gross Weight(KG) | 28 |